| Portable atmosphere testing instruments — enclosed space entry | Per manufacturer's schedule — commonly 6 or 12 months | Manufacturer's instructions. SOLAS Reg. XI-1/7 requires the instrument and requires suitable means for its calibration to be provided; it sets no interval. IMO Res. MSC.581(110) para. 7.2 recommends testing with properly calibrated equipment and that the manufacturers' instructions be strictly followed. | Oxygen, flammable gases or vapours, hydrogen sulphide and carbon monoxide as a minimum, in one instrument or several |
|---|
| Bump and response test, portable gas detectors | Before each use | Manufacturer's instructions and the vessel's SMS under the ISM Code | A bump test proves response, not accuracy. It is not a calibration |
|---|
| Calibration gas — cylinder shelf life | Per the cylinder's own expiry date | Gas manufacturer's certificate of analysis. Reactive mixtures — hydrogen sulphide in particular — carry a shorter shelf life than oxygen or methane mixtures. | Expired gas invalidates the span; the cylinder is named on the certificate with batch number and expiry date |
|---|
| Fixed gas detection and sampling systems | Per manufacturer's schedule | Manufacturer's instructions and the vessel's planned maintenance system under the ISM Code. No IMO instrument fixes an interval. | Sample lines and the alarm chain proved end to end |
|---|
| Pressure, vacuum and temperature instruments | Per PMS or class survey cycle | The vessel's SMS under the ISM Code, and classification society rules where the instrument forms part of a class-surveyed system | No IMO instrument sets a calibration interval for these |
|---|
| Tank gauging systems, UTI tapes, level instruments | Per manufacturer's schedule — commonly 12 months for UTI tapes | Manufacturer's instructions; classification society rules for the gauging installation | Inspectors and terminals read the date on the certificate and on the tape's label; both must agree |
|---|
| Loading computer and stability instrument verification | Annual | Classification society rules — check against the approved test conditions at each annual survey | Repeat after any software or approved-condition change |
|---|
| Oil content meter — 15 ppm bilge alarm | At intervals not exceeding 5 years after commissioning, or the term stated in the manufacturer's instructions, whichever is shorter | IMO Res. MEPC.107(49) para. 4.2.11, as replaced by MEPC.285(70). The accuracy check is by calibration and testing conducted by a manufacturer or persons authorised by the manufacturer; alternatively the unit is replaced by a calibrated one. | Certificate validity is checked at IOPP annual, intermediate and renewal surveys; the certificate is retained on board |
|---|
| ODME — oil discharge monitoring and control | Accuracy verified at each IOPP Certificate renewal survey — 5 years maximum | IMO Res. MEPC.108(49), as amended by MEPC.240(65), para. 5.3 — which applies to ODMEs installed in oil tankers whose keels were laid on or after 1 January 2005; equipment on earlier tankers may instead sit under A.393(X), A.496(XII), MEPC.13(19) or A.586(14). MARPOL Annex I Reg. 31 requires the system on oil tankers of 150 gross tonnage and above; the manufacturer's instructions govern any shorter interval. | Calibration certificate showing the date of the last calibration check retained on board |
|---|
| Flow meters and bunker measurement equipment | Per manufacturer's schedule | Manufacturer's instructions. Some ports impose their own verification regime under national metrology law. | MARPOL Annex VI requires bunker delivery notes, not meter calibration |
|---|
| Torque wrenches — hand torque tools | 12 months or 5,000 cycles, whichever occurs first | ISO 6789-2:2017 clause 4.1 — a default that applies where the user operates no control procedure; where one exists, the interval is set from the tool's conditions of use. ISO 6789 covers hand torque tools only. Not an IMO requirement. | Powered tools follow the manufacturer's instructions; any tool is recalibrated after overload, repair or improper handling |
|---|
| Load cells and dynamometers for load testing | Certificate current at the time of test | Class and flag requirements for test equipment used under SOLAS Reg. II-1/3-13 (in force 1 January 2026) and ILO Convention No. 152 | The interval is that of the test equipment's own certificate |
|---|
| Reference standards used on the job | Per each standard's own recalibration interval | Each reference standard carries its own calibration certificate from the laboratory that issued it, and that certificate — the issuing laboratory and its reference — is named on yours, so the chain behind a reading is checkable at its source. We do not publish our own accreditation status on a marketing page; what is held for a quoted scope is stated in writing with the quotation. | Named on the certificate with its own certificate reference |
|---|
| Recalibration after repair, sensor change or impact | Event-driven | Manufacturer's instructions and the vessel's SMS under the ISM Code | Also after a failed verification, before the instrument returns to service |
|---|